How to fit a C 1s XPS spectrum
The C 1s region shows up in almost every XPS measurement, and it is often where fitting goes wrong. This guide walks through a sensible, repeatable approach, from charge correction to checking the result.
1. Correct for charging first
Insulating and poorly grounded samples charge under the X-ray beam, which shifts every peak in the spectrum by the same amount. Before you fit anything, shift the energy axis so a known reference sits where it should. The most common choice is adventitious carbon (the thin layer of hydrocarbon contamination on most air-exposed samples), set to 284.8 eV.
Adventitious carbon is a convenient reference, not a perfect one. If your sample has its own carbon, or you use another reference such as a metal Fermi edge or a substrate peak, use that and say so when you report results.
2. Pick the region and the background
Select an energy window that starts and ends on flat baseline on both sides of the C 1s envelope. For most core levels a Shirley background is the standard choice. A linear background is fine for simple, well-separated peaks, and a Tougaard background is an alternative when the spectrum extends well beyond the peak on the high binding energy side.
3. Add components for the chemical states you expect
Each component corresponds to one chemical environment. Typical carbon states in organic and polymer samples sit at roughly these binding energies once the spectrum is charge corrected:
| Environment | Approximate binding energy |
|---|---|
| C–C, C–H | 284.8 eV (reference) |
| C–O (alcohols, ethers) | about 286.0–286.5 eV |
| C=O (carbonyl) | about 287.5–288.0 eV |
| O–C=O (carboxyl, ester) | about 288.5–289.5 eV |
These are typical values, not fixed ones. Exact positions depend on your sample, your reference, and the instrument, so check the literature for your system before assigning chemistry.
4. Constrain what physics tells you
A fit with every position, width and area free can find many equally good solutions that make no chemical sense. Constraints keep it honest:
- Equal widths. Components of the same element measured together usually have similar widths. Tie them with a constraint such as
A.fwhm, so the fit cannot use one component's width to swallow a neighbour. - Fixed separations. If you know two states are about a fixed distance apart, tie the position, for example
A.position + 1.5. - Sensible line shapes. A Gaussian-Lorentzian mix such as GL(30) is a common starting point for non-metallic carbon. Keep the shape parameters fixed unless your data clearly determine them.
5. Check the fit, not just the picture
- Look at the residual (data minus fit). Structure in the residual means a missing or mis-shaped component, or a background mismatch.
- With counting-noise weighting, a reduced χ² near 1 means the model describes the data to within noise. Values well above 1 point to a problem. R² alone cannot tell a good fit from a slightly wrong one.
- Compare models with AIC or BIC. Each extra component should lower BIC by a clear margin, otherwise it is probably fitting noise.
- Ask whether the result is chemically reasonable. A component that only exists to make the numbers fit is not evidence of a chemical state.
6. Read off the component shares
Once the fit is good, the area of each component gives the share of carbon in that chemical environment. Report these as percentages of the C 1s envelope, together with the fit uncertainty on each.
Common mistakes
- Fitting before correcting for charging, then assigning chemistry to shifted peaks.
- Adding components until the residual looks flat, with no chemical reason for them.
- Letting widths float independently so one component absorbs its neighbour.
- Choosing a region that is too narrow, so the background is poorly defined.
- Treating a single fit as the answer. It is a model; check it against the chemistry of your sample.
Doing this in XPSLab
- Import your file and select the C 1s scan.
- Press
Rto auto-detect the strongest peak and create a region with a Shirley background, or drag across the C 1s range yourself. - Use the C 1s ref charge-correction button to set adventitious carbon to 284.8 eV.
- Click Detect peaks & fit for a starting set of components, or add them yourself and click Refine fit.
- Add constraints in each component's constraint fields, then check the Fit Statistics panel: reduced χ², residual, and parameter correlations.
- Read each component's share in the Quantification table, which also gives a total row for the whole envelope.
The automatic fit is a starting point. Always check it against what you know about your sample.